This unit has been designed in accordance with Chapter 7 Section 2 Third Edition, May 2018.
FEATURES
- User selectable auto-off timer to save battery life
- Resolution of 0.1 or .01
- Temperature range -40 to 400 degrees F
- Averaging function
- Auto-Sample function with average feature
- F to C switching
- Backlit LCD with indicating arrows indicating rate of change
- 4 Point calibration
- Cable units available
- Anodized coating for corrosion protection
- Highly durable recessed membrane switch faceplate
- Glow in the dark keypad
- Auto backlight upon power up function (Hold)
SPECIFICATIONS
Display: Amber backlit LCD with 0.45" characters with trend indication arrows and low battery and fault indication
Construction: Stainless steel temperature probe and sensor. Electronics and display in 4" x 2" x 7" anodized Aluminum enclosure. Highly durable display window, graphics and membrane switches.
Weight: Weight depends on cable length and probe style. This unit employs a heavier probe ideal for use in heavier products such as asphalt.
Temperature: Resolution to 0.01 Degrees
Calibrated Accuracy: +-0.080° F from 32° to 300° F +-0.048° C from 0° to 150° C
Time Constant: 7 seconds in circulated water bath Constant: (63.2% response to temperature change)
Sensor: 100 Ohm, Class A Platinum RTD Immersion depth 4", 11cm
Battery: Energizer Ultimate Lithium L91 Intrinsically Safe Rated AA, Battery life (depending upon use) 25-150 hrs Battery Operating Range: -40°C to 60°C (-40°F to 140°F)
TYPICAL APPLICATIONS: Custody transfer, temperature baths, reference, temperature calibration for flow meters, thermowells, provers, reference, distillations, melting point tests, and flash point tests.
Industries: Laboratory, Petroleum, Pipeline, Pharmaceutical, Food & many more.
Certifications: Intrinsically safe for use in Class I, Division I Groups A, B, C, & D. Class 1, Zone 0, Group IIC Hazardous Locations CSA C/US A/Ex ia IIC T4